Product specifications, 1. list of specifications, Common specifications – RKI Instruments RM-5000 Series Indicator/Alarm Unit User Manual
Page 77

10 Product Specifications
10-1. List of specifications
- 77 -
RM-5000
10
Product Specifications
10-1. List of specifications
Concentration display Character LCD (digital and bar meter display
red>)
Power display
POWER lamp on (green)
Gas alarm display
First: ALM1 lamp blinks or lights up (red)/Second: ALM2 lamp blinks or lights up
(red)
Gas alarm pattern
Self-latching, auto-reset, or lock-in
Gas alarm contact
No-voltage contact 1a or 1b (2 step independent)
De-energized (energized at an alarm) or energized (de-energized at an alarm)
Fault alarm/self
diagnosis
System abnormalities/detection circuit abnormalities/communication
abnormalities (digital transmission only)
Fault alarm display
FAULT lamp blinking (orange)/Detail display
Fault alarm pattern
Auto-reset
Fault alarm contact
No-voltage contact 1a or 1b
De-energized (energized at an alarm) or energized (de-energized at an alarm)
Contact capacity
100 VAC - 0.5A/30 VDC - 1.5A (resistant load)
Transmission system
Analog/digital transmission [option]
Transmission
specifications
Analog transmission: 4 - 20 mA DC (no-insulation/load resistance under 300
Ω)/Digital transmission: RS-485
Power supply
24 VDC (21.6 – 26.4 VDC)
Initial clear
Approx. 25 seconds
Operating
temperatures
-10 – 40°C (at a constant condition)
Operating humidities
10 to 90%RH (Non-condensing)
Structure
Card type with front display used enclosed in a case (a single-unit or multi-unit
case)
External dimensions
Approx. 29.6 (W) x 120 (H) x 92 (D) mm (projection portions excluded)
Weight
Approx. 0.10 kg (unit only)
* Specifications subject to changes without notice.