7 Maintenance

7-3. Gas calibration method

RM-5000

- 70 -

This is used to perform the span adjustment on the detector head (sensor). For the oxygen deficiency alarm
specification (O2: 0 - 25 vol%), this is the same as “1-1”.
<>

2-2. SPAN

Press SET key.

Gas Introduction

Introduce a gas to the connected detector
head (sensor), and then press the SET
key when the reading is stabilized.
* In case of less than 10%FS, this is not

switched to the next menu.

Adjustment Result Recording

To record the adjustment result, press the
SET key.
(Press the MODE key to cancel this
menu.)

Span Adjustment

Because the reading will be fixed, adjust
the introduced gas concentration by
pressing the ▲ or ▼ key. After adjusting
the value, press the SET key.
(20.9 vol% in the oxygen deficiency alarm
specification)

During maintenance
During inhibit

Lamp on
Lamp off
Lamp blinking

Span Adjustment Completed

Adjustment Completed

The menu returns to

2-2. SPAN automatically.

* When the span adjustment fails

Return to 2-2. SPAN

Return to 2-2. SPAN


Do not press the SET
key before the reading
is stabilized.


Exhaust gas must be collected in
the exhaust bag or discharged
through the exhaust line.