7 Maintenance
7-2. Regular maintenance mode
RM-5000
- 58 -
<
During maintenance
During inhibit
Lamp on
Lamp off
Lamp blinking
SE 1. ALM P
Press SET key.
Return to SET 1.ALM P
First Alarm Setpoint Value
Setting
Press the ▲ or ▼ key to change
the value, and then press the
SET key to confirm the value.
Second Alarm Setpoint Value
Setting
Press the ▲ or ▼ key to change
the value, and then press the
SET key to confirm the value.
During maintenance
During inhibit
Lamp on
Lamp off
Lamp blinking
SE 3.F TEST
Press SET key.
Fault Test ON/OFF
Select either ON/OFF.
Press the ▲ or ▼ key to select ON in
the display, and then press the SET
key to activate the fault alarm.
Select OFF in the display and press
the SET key to deactivate this
menu.(Alternatively, press the MODE
key to deactivate this menu.)
WARNING
Because the contact (fault) can be activated only by a fault alarm test in the maintenance mode, be
careful to perform the test. The fault alarm test cannot be performed during inhibit ( I ).