3. other adjustments/cleaning method, Flow, 50 l/m – RKI Instruments GD-70D Series User Manual
Page 67: Def flow

8 Maintenance
8-3. Other adjustments/Cleaning method
- 67 -
GD-70D
8-3. Other adjustments/Cleaning method
The flow rate of the detector is automatically adjusted to 0.5 L/min. Turning off the auto-adjustment function
enables the manual adjustment. (See 2-10 - SET-11)
The manual flow rate adjustment can be performed in the regular maintenance mode "2-6.FLOW".
PW
A1
A2
F LCD
2-6.FLOW
Press the SET key.
●
○
○
○
2-6
FLOW
MAINTENANCE
↓
↓
●
○
○
○
50 %
0.50 L/M
MAINTENANCE
Flow Rate Adjustment
As the output is increased or decreased by
pressing the ▲ or ▼ key, the flow rate
changes accordingly. Adjust the flow rate to
0.5 L/min, and then press the SET key.
(Press the MODE key to cancel this menu.)
Regardless of the flow rate auto-adjustment or manual-adjustment, when the reading on the flow rate
indicator is incorrect (due to aging degradation, improper flow rate default set, or other reasons), it must be
set to provide the right flow rate.
To set the Flow Rate Indicator, prepare and connect a flow meter (which indicates 0.5 L/min precisely), and
perform the default set with the flow meter while it is indicating 0.5 L/min in the regular maintenance mode
"2-5.DEF FLOW".
PW
A1 A2 F
LCD
2-5.DEF FLOW
Press the SET key.
●
○
○
○
2- 5
DEF FLOW
MAINTENANCE
↓
↓
●
○
○
○
1000
DEF FLOW
MAINTENANCE
Flow Rate Adjustment/Default Set
Adjust the flow rate by pressing the ▲ or ▼
key until the connected flow meter indicates
0.5 L/min
± 10%, and then press the SET
key.
(Press the MODE key to cancel this menu.)
CAUTION
Perform the flow rate default set after verifying that the suction volume is 0.5 L/min by using a flow meter.
WARNING
After the adjustment is completed, do not forget to press the MODE key to return to the detection mode.