Set 15, Hold, 20ma – RKI Instruments GD-70D Series User Manual
Page 58: Set 16, Set 17, Save, Set 18, Mnt out, Ma 4-20, 4ma adj

7 Regular Maintenance Mode
GD-70D
- 58 -
▲
↓ ↑
▼
● ○ ○
○
SET 15
MNT OUT
MAINTENANCE
⇔
SET
2.5mA
MNT OUT
MAINTENANCE
▲
↓↑
▼
4.0mA
MNT OUT
MAINTENANCE
▲
↓↑
▼
HOLD
MNT OUT
MAINTENANCE
▲
↓↑
▼
4-20mA
MNT OUT
MAINTENANCE
SET 15. MNT OUT
Set the external output for the
maintenance mode.
Select either 2.5 mA/4.0
mA/HOLD (previous value)/4
- 20 mA (linked to display
value), and then press the
SET key to confirm the
selection.
(4 - 20 mA specification)
| ↑
| |
| |
| |
▲
| |
▼
| |
| |
| |
↓ |
● ○ ○
○
SET 16
MA 4-20
MAINTENANCE
⇔
SET
100%
4MA ADJ
MAINTENANCE
↓
100%
20MA ADJ
MAINTENANCE
SET 16.MA 4-20
Adjust the external output (4 -
20 mA).
Adjust the output (%) by
pressing the ▲ or ▼ key, and
then press the SET key to
confirm the value.
(It must be adjusted to the
upper unit or the ammeter.)
After adjustment of 4 mA is
completed, perform the
adjustment of 20 mA.
(Press the MODE key to skip
this menu.)
(4 - 20 mA specification)
| ↑
| |
▲
| |
▼
| |
↓ |
● ○ ○
○
SET 17
BK LIGHT
MAINTENANCE
⇔
SET
ON
BK LIGHT
MAINTENANCE
▲
↓↑
▼
SAVE
BK LIGHT
MAINTENANCE
SET 17.BK LIGHT
Set the backlight.
Select either ON/SAVE, and
then press the SET key to
confirm the selection.
When ON is selected, the
backlight lights up
continuously.
When SAVE is selected, the
backlight is usually turned off,
but lights up only during an
operation or event.
| ↑
▲
| |
▼
↓ |
● ○ ○
○
SET 18
ETHERNET
MAINTENANCE
SET 18. ETHERNET
This is a setting screen of the
ETHERNET setting. This is
not used for the detector.
▲
↓ ↑
▼