Set 6, Rly ptrn, Al1 rly – RKI Instruments GD-70D Series User Manual
Page 61: Al2 rly, Flt rly

7 Regular Maintenance Mode
- 61 -
GD-70D
Energized/De-Energized Contact Setting
PW A1 A2
F
LCD
SET 6. RLY PTRN
Press the SET key.
● ○ ○ ○
SET 6
RLY PTRN
MAINTENANCE
↓
↓
● ○ ○ ○
nd
AL1 RLY
MAINTENANCE
⇔
▲/▼
nE
AL1 RLY
MAINTENANCE
First Alarm Contact
Setting
Select either
nd(De-Energized)/nE(Energi
zed), and then press the
SET key to confirm the
selection. (Press the MODE
key to skip this menu.)
↓
↓
● ○ ○ ○
nd
AL2 RLY
MAINTENANCE
⇔
▲/▼
nE
AL2 RLY
MAINTENANCE
Second Alarm Contact
Setting
Select either nd
(De-Energized)/nE
(Energized), and then press
the SET key to confirm the
selection. (Press the MODE
key to skip this menu.)
↓
↓
● ○ ○ ○
nd
FLT RLY
MAINTENANCE
⇔
▲/▼
nE
FLT RLY
MAINTENANCE
Fault Alarm Contact
Setting
Select either nd
(De-Energized)/nE
(Energized), and then press
the SET key to confirm the
selection. (Press the MODE
key to skip this menu.)
↓
Return
to
SET 6. RLY PTRN
NOTE
When De-energized is selected, the relay is energized and activated in response to an alarm (de-energized
at a normal environment).
• When the contact "a" is used, it is open at a normal environment while closed in response to an alarm.
• When the contact "b" is used, it is activated conversely.
When Energized is selected, the relay is energized at a normal environment (de-energized in response to
an alarm).
• When the contact "a" is used, it is closed at a normal environment while open in response to an alarm.
In addition, it is closed when the power is OFF.
• When the contact "b" is used, it is activated conversely.