2ppm, 0 mv, M mode – RKI Instruments GD-70D Series User Manual
Page 50: Gas test, Sih4, Zero, Span, Last cal, Bias

7 Regular Maintenance Mode
GD-70D
- 50 -
Regular Maintenance Mode
PW
A1
A2
F
LCD
User Mode
In "1-8.M MODE", press the
SET key.
● ○ ○
○
1- 8
M MODE
MAINTENANCE
↓
↓
Then press the SET key again
for three seconds.
● ○ ○
○
- - - -
M MODE
MAINTENANCE
↓
Regular Maintenance
Mode
↓
● ○ ○
○
2- 0
GAS TEST
MAINTENANCE
→
SET
0.2ppm
GAS TEST
MAINTENANCE
↓
(Alternate Display)↑
0.2ppm
SIH4
MAINTENANCE
2-0.GAS TEST
Perform a test with the gas.
Similar to the detection mode,
the reading changes and the
alarm lamp lights up after the
gas is introduced, but the
contact is not activated.
| ↑
▲
| |
▼
↓ |
● ○ ○
○
2- 1
ZERO
MAINTENANCE
→
SET
Zero Adjustment
⇒P65
2-1.ZERO
Perform the zero adjustment.
▲
↓ ↑
▼
● ○ ○
○
2- 2
SPAN
MAINTENANCE
→
SET
Span Adjustment
⇒P66
2-2.SPAN
Perform the span adjustment.
▲
↓ ↑
▼
● ○ ○
○
2- 3
LAST CAL
MAINTENANCE
→
SET
12:00
2009.01.01
MAINTENANCE
2-3.LAST CAL
Show the last calibrated date.
▲
↓ ↑
▼
● ○ ○
○
2- 4
BIAS
MAINTENANCE
→
SET
0 mV
BIAS
MAINTENANCE
2-4.BIAS
Show the bias voltage "BIAS"
of electrochemical type (ESU)
or the element voltage
"EVOLT" of semiconductor
type (SGU).
▲
↓ ↑
▼
PW: POWER
A1: ALM1
A2: ALM2
F: FAULT
●: Lamp on