Set 10, Set 11, Set 12 – RKI Instruments GD-70D Series User Manual
Page 57: Set 13, Set 14, Flt ptrn, At flow, Zero f, Zero 24f, S assist

7 Regular Maintenance Mode
- 57 -
GD-70D
▲
↓ ↑
▼
● ○ ○
○
SET 10
FLT PTRN
MAINTENANCE
SET 10. FLT PTRN
This is a setting screen of the
fault alarm activation. Do not
change the setting when the
detector is used in a normal
way, because it determines
how the detector functions.
(Auto-reset setting: "nL")
▲
↓ ↑
▼
● ○ ○
○
SET 11
AT FLOW
MAINTENANCE
⇔
SET
ON
AT FLOW
MAINTENANCE
▲
↓↑
▼
OFF
AT FLOW
MAINTENANCE
SET 11.AT FLOW
Set the flow rate
auto-adjustment.
Select either ON/OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
flow rate auto-adjustment is
activated.
| ↑
▲
| |
▼
↓ |
● ○ ○
○
SET 12
ZERO F
MAINTENANCE
⇔
SET
ON
ZERO F
MAINTENANCE
▲
↓↑
▼
OFF
ZERO F
MAINTENANCE
SET 12. ZERO F
Set the zero follower when
the electrochemical type
(ESU) or the
pyrolysis-particle type (SSU)
is installed.
Select either ON/OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
zero follower function is
activated.
* When SSU is installed, you
must select ON.
| ↑
▲
| |
▼
↓ |
● ○ ○
○
SET 13
ZERO 24F
MAINTENANCE
SET 13.ZERO 24F
This is a supplemental setting
screen for the above zero
follower function. (A setting to
determine whether the
first-24-hour zero follower will
be performed after the power
is turned on.)
Do not change the setting
when the detector is used in a
normal way.
(ON setting)
| ↑
▲
| |
▼
↓ |
● ○ ○
○
SET 14
S ASSIST
MAINTENANCE
⇔
SET
ON
S ASSIST
MAINTENANCE
▲
↓↑
▼
SET 14.S ASSIST
Set the sensitivity correction
when the electrochemical
type (ESU) is installed.
Select either ON/OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
sensitivity correction function
is activated.
| ↑
▲
| |
▼
↓ |
OFF
S ASSIST
MAINTENANCE