0 °c, Def flow, Flow – RKI Instruments GD-70D Series User Manual
Page 51: 50 l/m, Temp, Warmtime, Setting1, Setting2, Pl data

7 Regular Maintenance Mode
- 51 -
GD-70D
▲
↓ ↑
▼
● ○ ○
○
2- 5
DEF FLOW
MAINTENANCE
⇔
SET
1000
DEF FLOW
MAINTENANCE
2-5.DEF FLOW
Set the flow rate to the default
value when the flow sensor is
replaced, etc.
Normally, this is not used
because the detector has
already been adjusted.
On the other hand, if the flow
meter is set at other than the
specified flow rate, gases
cannot be detected properly
due to an incorrect flow rate.
(Press the MODE key to
cancel this menu.)
| ↑
| |
▲
| |
▼
↓ |
● ○ ○
○
2- 6
FLOW
MAINTENANCE
⇔
SET
50 %
0.50 L/M
MAINTENANCE
2-6.FLOW
Show the output and flow rate
of the current pump.
▲
↓ ↑
▼
● ○ ○
○
2- 7
TEMP
MAINTENANCE
→
SET
25.0
°C
TEMP
MAINTENANCE
2-7.TEMP
Show the temperature of the
detector.
▲
↓ ↑
▼
● ○ ○
○
2- 8
WARMTIME
MAINTENANCE
→
SET
12:00
2009.01.01
MAINTENANCE
2-8. WARMTIME
Show the suggested warm-up
completion date/time for
semiconductor type (SGU),
which requires a long
warm-up time.
▲
↓ ↑
▼
● ○ ○
○
2- 9
SETTING1
MAINTENANCE
→
SET
Environmental
Setting 1
⇒P53
2-9. SETTING1
Specify the environmental
setting 1.
▲
↓ ↑
▼
● ○ ○
○
2-10
SETTING2
MAINTENANCE
→
SET
Environmental
Setting 2
⇒P55
2-10.SETTING2
Specify the environmental
setting 2.
▲
↓ ↑
▼
● ○ ○
○
2-11
PL DATA
MAINTENANCE
→
SET
For information on
Pyrolyzer Data
Display, see
"PLU-70 Operating
Manual".
2-11.PL DATA
Show the pyrolyzer data when
the pyrolyzer unit (PLU-70) is
used.
▲
↓ ↑
▼