Set 0, Set 1, Set 2 – RKI Instruments GD-70D Series User Manual
Page 55: 9ppm, Set 3, Slope, Set 4, Setting2, Address, Day time

7 Regular Maintenance Mode
- 55 -
GD-70D
In the environmental setting 2, specify the settings of functions. (* It is recommended that setting changes
should be recorded in a log.)
The environmental setting 2 includes setting menus which are usually not used. Be careful not to change
these settings by mistake.
Environmental Setting 2
PW
A1
A2
F
LCD
2-10.SETTING2
Press the SET key.
● ○ ○
○
2-10
SETTING2
MAINTENANCE
↓
↓
● ○ ○
○
SET 0
ADDRESS
MAINTENANCE
⇔
SET
01
ADDRESS
MAINTENANCE
SET 0. ADDRESS
Set the address.
(NT specification)
Select the number by
pressing the ▲ or ▼ key, and
then press the SET key to
confirm the address.
▲
↓ ↑
▼
● ○ ○
○
SET 1
DAY TIME
MAINTENANCE
⇔
SET
Date/Time Setting
⇒P60
SET 1.DAY TIME
Set the date/time of the main
unit.
▲
↓ ↑
▼
● ○ ○
○
SET 2
SUPPRESS
MAINTENANCE
⇔
SET
0.9ppm
SUPPRESS
MAINTENANCE
SET 2.SUPPRESS
Set the suppression value.
Select the number by
pressing the ▲ or ▼ key, and
then press the SET key to
confirm the value.
▲
↓ ↑
▼
● ○ ○
○
SET 3
SUP TYPE
MAINTENANCE
⇔
SET
CUT
SUP TYPE
MAINTENANCE
▲
↓↑
▼
SLOPE
SUP TYPE
MAINTENANCE
SET 3. SUP TYPE
Set the suppression system.
Select either CUT/SLOPE,
and then press the SET key
to confirm the selection.
When CUT (Cut Off) is
selected, values exceeding
the suppression value are
directly displayed. When
SLOPE (Slope) is selected,
values exceeding the
suppression value are gently
displayed.
| ↑
| |
▲
| |
▼
↓ |
● ○ ○
○
SET 4
TEST RLY
MAINTENANCE
⇔
SET
OFF
TEST RLY
MAINTENANCE
▲
↓↑
▼
SET 4. TEST RLY
Set the contact activation for
an alarm test.
Select either ON/OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
contact can be activated even
during an alarm test.
| ↑
▲
| |
▼
↓ |
ON
TEST RLY
MAINTENANCE