2 Regular Maintenance Mode

- 9 -

GX-8000

<>

The maintenance mode includes setting menus which are usually not used. Be careful not to change these
settings by mistake. It is recommended that setting changes should be recorded in a log.


Keep the S and T switches pressed and
press the POWER switch. When the buzzer
beeps, release the switches.

The input field blinks. Press the S or T switch
to select a number and press ENTER to
confirm it. Enter four digits from the left.
Password: 0008

Regular maintenance mode

DATE
Set the date and time.

Date/time setting

=> P11

AIR CAL

Perform the fresh air adjustment.

Fresh air
adjustment

=> P16

VOL Z.CAL

Perform the zero adjustment for
high-concentration combustible gases.

Zero adjustment
for high-
concentration
combustible
gases

=> P20

AUTO CAL

Perform the span adjustment simultaneously
for all channels.
(Set the span gas concentration in advance
and make adjustment at once.)

Simultaneous
span adjustment
for all channels

=> P18

ONE CAL

Perform the span adjustment for each channel.
(While gas is supplied, perform span
adjustment using UP/DOWN.)

Span adjustment
for each channel

=> P19