2 Regular Maintenance Mode

- 5 -

GX-8000

<>


Fresh air
adjustment

=> P16

In the detection mode, keep T pressed and press
the DISPLAY switch. When the buzzer beeps,
release the switches.

Calibration mode
AIR CAL

Perform the fresh air adjustment.

Zero adjustment
for high-
concentration
combustible
gases

=> P20

VOLZ.CAL

Perform the zero adjustment for
high-concentration combustible gases.

Simultaneous
span adjustment
for all channels

=> P18

AUTO CAL

Perform the span adjustment simultaneously for all
channels.
(Set the span gas concentration in advance and
make adjustment at once.)

Span adjustment
for each channel

=> P19

ONE CAL

Perform the span adjustment for each channel.
(While gas is supplied, perform span adjustment
using UP/DOWN.)

Bump test

=> P6

BUMP

Perform the bump test.

Detection mode

NORMAL

Exit the calibration mode.