beautypg.com

Span adjustment, 1. preparation for span adjustment, Warning – RKI Instruments GX-8000(H4-0050) User Manual

Page 14: Caution

background image

3 Span Adjustment

3-1. Preparation for span adjustment

GX-8000

- 14 -





3

Span Adjustment




3-1. Preparation for span adjustment

Required equipment/material

• Zero adjustment gas for high-concentration combustible gases (collected in a gas sampling bag)

• Span adjustment gas (collected in a gas sampling bag)

• Stopwatch


Connection

Connect the equipment as shown below to perform the span adjustment.






























Gas

sampling

bag

Gas sampling probe

(with a tube)

GX-8000
main unit

GAS IN

GAS OUT

(Open)

WARNING

About the span adjustment gas
The span adjustment gas is a hazardous gas (toxic, oxygen deficient, etc.). Handle the gas and
related jigs and tools with due care (e.g., the gas must not be inhaled or the gas sampling bag must
not have any hole).

About the place for span adjustment

• Perform span adjustment where no silicon, organic solvent, spray can gases, etc. are used.
• Perform span adjustment indoors at normal temperature without remarkable fluctuation (within ±5°C).
• Perform span adjustment in an exhaust booth.

CAUTION

The GAS OUT side of the pipe must be left open without any pipe connected. A supplied gas must be
discharged to a safe place.