2 Regular Maintenance Mode

GX-8000

- 8 -

NOTE

If "F" (Fail) is displayed after the span adjustment, replace the sensor with a new one.



Return to BUMP.

(2)

Continue
supplying the
test gas.

Stop supplying
the test gas.


Continue supplying the gas. The result is not
displayed, and the span adjustment is
automatically started. CAL and APPLY are
displayed alternately, and the countdown is
started.

The span adjustment is started in about 30
seconds, and the result is displayed. Left: Test
result, Right: Adjustment result (P: Pass, F:
Fail)
The results (P or F) <-> bump test result
values <-> span adjustment result values are
displayed alternately every time the S or T
switch is pressed. Stop supplying the gas.

After checking the result, press ENTER. The
gas monitor returns to the calibration mode
menu.