Retsch SK 100 User Manual
Page 11

05/04/2012
Retsch GmbH
11
0002
Setting up on a laboratory table
Fig.3
You can mount the SK 100 on your laboratory table. Fig. 3
*
Measure laboratory table thickness
*
Use screws with a maximum diameter of 8mm and a
length of the laboratory table thickness + 25 mm
*
Stand the SK 100 including collecting vessel on the
table
*
Mark the holes for the SK 100 for the table with ref-
erence to the holes on the motor
The distance between the holes on the front edge of the table
must be such that the collecting vessel or filter hose can be
mounted and dismounted easily.
Electrical connection
*
The voltage and frequency of the SK 100 can be found
on the nameplate.
*
Ensure that the values agree with the power supply avail-
able.
*
Connect the SK 100 to the power supply using the con-
necting cable supplied.
The power cable supplied does not have a plug because the
type of plug depends on the setting up location and the respec-
tive national regulation.
* When connecting the power cable to the power supply,
provide an external fuse in accordance with the regulations
of the setting-up location.
If the values are not observed, electronic and mechanical compo-
nents may be damaged.
Before first use, the direction of rotation must be checked, see arrow
on motor.
If the direction of rotation is incorrect, the grinding will be insuffi-
cient and mechanical components may be damaged.