Operating the machine, 4 operating elements and displays – Retsch Cryomill User Manual
Page 29
Operating the machine
29
Pos : 6. 11 /00005 Ü berschrif ten/ 1. 1 Ü bersc hrift en/ 1. 1 Übersc hrift en BD A/ 11 Bedienel ement e und Anzeig en @ 0\ mod_1226566557086_9. doc @ 3543 @ @ 1
5.4 Operating elements and displays
Pos : 6. 12 /00010 Bedi enungsanleit ung en Kapitels ammlungen/Cr yoMill II /0007 Bedi enung/0002 Cr yoMill II Modul Anz eige und Bedi en einheit @ 5\ mod_1343829230960_9.doc @ 33158 @ @ 1
Fig. 19: Control unit
Element
Description
Function
B1
LED on
– valve open
LED off
– valve closed
Indicates whether the internal solenoid valve on the
liquid nitrogen supply line is open or closed.
B2
LED on
– liquid nitrogen is flowing
LED off
– no liquid nitrogen
Indicates that liquid nitrogen is flowing into the device.
B3
LED precooling is running (Vkz)
Precooling is taking place / lights up when values are
set
B4
LED grinding is running (Mz)
The sample is ground / lights up when values are set
B5
LED intermediate cooling is running
(Zkz)
Intermediate cooling is taking plate / lights up when
values are set
B7
Frequency minus button
In setting mode and in operating mode:
reduces the set frequency by 1 Hz (1/s). Continuous
pressing switches on fast scrolling of numbers.
Frequency plus button
In setting mode and in operating mode:
increases the set frequency by 1 Hz (1/s). Continuous
pressing switches on fast scrolling of numbers.
B8
Cooling cycle minus button
Reduces the number of total cycles by 1
Cooling cycle plus button
Increases the number of total cycles by 1
B9
Time minus button
In setting mode and in operating mode:
Reducing the set
•
Precooling
•
Grinding time
•
Intermediate cooling time
B10
B7
B8
B15
B2
B1
B11
B12
B13
B9
B14
B3
B4
B5