RaySafe Solo User Manual
Page 42

The RaySafe Solo detector end trigs during exposure.
Make sure to select an appropriate Calc. delay if using the instrument with pulsed generators, or with
generators having a pre-pulse before the main exposure.
The RaySafe Solo does not respond when pressing keys.
try the reset switch and make sure that the battery is charged.
I get strange readings when measuring HVL on mammography, in comparison with my manual HVL
measurements.
Please make sure that the thickness of your Al filters is accurately known. A resolution on the order of 1 µm
is needed for 0.1 mm thickness.
I get strange mA/mAs readings when I’m using the RaySafe Solo’s AC adaptor or when my computer is
connected to mains power.
Make sure that the instrument
• was connected to the generator during zero adjust.
• is operating on battery (no AC adaptor).
• is not connected to a grounded computer (use Bluetooth or a laptop running on batteries with no ethernet
cable connected).
42
RaySafe Solo User Manual – tips and tricks