1. SENSOR MENU
CT Dose
Ion chamber sensor
press STEP to step between sensors and SELECT to select a sensor.
NOTE!
When turning on the base unit with the RaySafe Solo CT detector connected, “Stabilizing” will
be displayed for a few seconds while the electronics are stabilizing.
2. MEASURE MODE
After an offset adjustment, the instrument is in MEASURE MODE. the displayed value is updated after each
exposure.
Dose
Gy (Air kerma, free in air) or R (exposure)
Dose Length Product Gycm or Rcm
NOTE!
The RaySafe Solo CT detector can be used in two dose modes (dose or dose length product
(DLP)), selectable in the
SETUP MENU. Since the RaySafe Solo CT chamber is 10.0 cm long,
DLP readings (in Gy•cm or R•cm) will always be exactly 10 times higher than dose readings.
3. SETUP MENU
press SELECT to enter the SETUP MENU (from MEASURE MODE) and STEP to step between setup
parameters.
All values are stored in a non-volatile memory and are valid until manually changed.
(factory settings in bold.)
Dose unit
Change the default units and/or use this feature to recalculate a value to a different
unit. (dose: Gy, Gycm, R or Rcm, temperature: °f or °C, pressure: kpa or mmhg)