MAMMoGRAphy MeASUReMent
Place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). Make 
sure that the center of the MAM sensor square is placed 6 cm from the examination table front edge.
1. SENSOR MENU
Mammography
Sensor for mammography measurements with submenu for 
selection of beam qualities, see list to the right (depending 
on configuration). The choice of “paddle” or “no paddle” only 
applies if the kV measurement is going to be made with or 
without a paddle in the beam and does not affect dose or HVL 
measurements.
Mo/Mo No Paddle 
Mo/Mo Paddle 
Mo/Rh 
Mo/Al 
Rh/Rh 
Rh/Al 
W/Rh No Paddle 
W/Rh Paddle 
W/Rh* No Paddle 
W/Rh* Paddle 
W/Ag
NOTE!
Select W/Rh* for measurements on Hologic Selenia and Planmed Nuance.
2. MEASURE MODE
The displayed values are updated after each exposure.
kV
kV (Mo/Mo, W/Rh)
(not available on RaySafe Solo MAM dose.)
Dose
Gy (Air kerma, free in air) or R (Exposure)