MAS: MEASUREMENT PARAMETER DEFINITIONS
mAs
is calculated from all recorded data.
mA
is calculated as an average of all samples above 50% of peak. Rush currents are automatically
removed. Intermediate readings are based on samples since last reading. For long measurements, the final
reading is recorded approximately 1–2 s before end trig.
Time
starts the first time the current reaches 50% of the calculated mA and ends the last time it drops
below 50%. Intermediate readings are time since start trig.
Pulses
are counted every time the signal reaches trig.
Pulse rate
and mAs per pulse are averages for the last 6 pulses.
Note!
If measurements are made simultaneously with a sensor and mAs cable, any shared parameter
(time, pulses or pulse rate) will be taken from the sensor.