MAS: MEASUREMENT PARAMETER DEFINITIONS

mAs

is calculated from all recorded data.

mA

is calculated as an average of all samples above 50% of peak. Rush currents are automatically

removed. Intermediate readings are based on samples since last reading. For long measurements, the final

reading is recorded approximately 1–2 s before end trig.
Time

starts the first time the current reaches 50% of the calculated mA and ends the last time it drops

below 50%. Intermediate readings are time since start trig.
Pulses

are counted every time the signal reaches trig.

Pulse rate

and mAs per pulse are averages for the last 6 pulses.

Note!

If measurements are made simultaneously with a sensor and mAs cable, any shared parameter

(time, pulses or pulse rate) will be taken from the sensor.

RaySafe X2 User Manual