MAM SENSOR: MEASUREMENT PARAMETER

DEFINITIONS

Dose

and HVL are calculated from all recorded data.

Dose rate

is the average dose rate, calculated as dose/time.

Time

starts the first time the dose rate waveform reaches 50% of peak, and ends the last time it drops

below 50%. Intermediate values are time since start trig.
kVp

is calculated from an average of samples above ~33 % of peak signal level.

Pulses

are counted at every start trig with more than 4 ms since last end trig.

Pulse rate

and dose per pulse are averages from the last 6 pulses.

Intermediate readings of dose rate, kVp and HVL are moving averages. Final readings are based on the
whole measurement.

Note!

If measurements are made simultaneously with the MAM sensor and mAs cable, any shared

parameter (time, pulses or pulse rate) will be taken from the MAM sensor.