MTS Triaxial Cell Models 656-06 User Manual
Page 68
Remove the Force Transducer
Note: Inspect and clean the shoulder bolt’s threads. Then lightly lubricate them with clean
confining fluid. Also lubricate all dry O-rings with clean confining fluid.
5. Install the O-rings as shown in this figure. Tighten the shoulder bolt to 20 lb•ft (28 N•m).
Install the Other Force Transducer
Description
Item
O-Ring
1
Tighten to 20 lb•ft (28 N•m)
2
Caution
:
Install the insert so the side with three holes faces out. If this side faces in, the insert’s
O-ring may leak.
6. Install the O-ring on the force transducer’s insert. Then use a nylon mallet to tap the insert in place.
68 | Triaxial Cell Product Information
Operation
This manual is related to the following products: