MTS Triaxial Cell Models 656-06 User Manual

Page 57

background image

Drain the Cell

The cell is drained when the confining intensifier’s reservoir begins bubbling vigorously.

Warning

:

You could get sprayed with debris and confining fluid if you open a Triaxial Cell that still
has pressure.

You could be hurt and your Triaxial Cell damaged.

Before opening the Triaxial Cell, make sure the Pressure Intensifier’s Confining Fluid Output
valve is Open. The Triaxial Cell is still pressurized if its bell starts rising as you undo the
last cap screws.

9. Vent any compressed air trapped inside the Triaxial Cell.

Vent the Cell

Triaxial Cell Product Information | 57

Operation