Calibration

27

B. Adjust the calibrator for 80% tension (or for the maximum

positive travel) of the range you are calibrating.

If the clip-on gage signal is above +8 V DC (or your
calculated output), adjust the

ΔK (delta K) control for a clip-

on gage signal of +8 V DC.

If the clip-on gage signal is below +8 V DC (or your
calculated output), the

ΔK adjustment cannot be made. Return

the

ΔK adjustment to its original setting and use the gain

control to increase the voltage halfway from its present value
to +8 V DC (this splits the difference between compression
and tension).

This completes the initial gain adjustment procedure using a calibration
stand.