How to use a calibration block, How to use a calibration block 23 – MTS Clip-On Gages User Manual
Page 23

Calibration
23
The following table shows the proper conditioner output for each clip-
on gage displacement of the preceding example.
.
How to Use a Calibration Block
MTS recommends using a calibration block to establish the initial gain
setting of the conditioner. The calibration block is a metal block with
knife edges mounted to it or machined in it. Using multiple knife edges
allows you to adjust the gain for a variety of specific displacements.
Use this procedure as a guideline to calibrate other clip-on gages.
1. Set up the controller.
Your test controller must be configured to use the clip-on gage
signal.
A. Connect the clip-on gage to your controller.
B. Monitor the excitation voltage.
C. Monitor the clip-on gage signal.
D. Prepare to use the calibration controls on your controller.
E.
Adjust the excitation control to set the excitation to +6 V DC
or the voltage specified in the clip-on gage documentation.
2. Zero the conditioner output.
% F
ULL
S
CALE
T
RAVEL
O
UTPUT
100%
3 mm
+10 V DC
80%
2.4 mm
+8 V DC
60%
1.8 mm
+6 V DC
40%
1.2 mm
+4 V DC
20%
0.6 mm
+2 V DC
Zero
0 mm
0 V DC
-20%
-0.6 mm
-2 V DC