MTS Insight - 1kN to 300kN User Manual

Page 48

background image

Insight Material Testing System

Controller Connections

Installation

48

Transformer Requirements for MTS Insight 150, 200 and 300 Machines

Important

Remember to verify local codes, voltages and frequencies
before ordering transformer.

Following are the minimum requirements for transformers for the MTS
Insight 150, 200, and 300 load frames.

Isolation Type Transformer (Electrical isolation between input and
output).

3-phase input and output.

Wye output with neutral connection at 400 VAC.

Equi-potential between each output phase and earth/ground.

Ambient operating temperature from 5 deg C to 40 deg C.

Temperature rise 150 deg C maximum.

Insulation class 200 deg C minimum.

Electrostatic Shield.

UL Listed along with CE and CSA/CUL or equivalent depending
country.

Additional

considerations

Floor standing.

NEMA 2 or better enclosure.

Harmonic Factor can be 0.

Maximum sound level 45 dB per ANSI Standard C89.2.

Typical Power in

Selected Areas

United States - Voltage and Frequency

480 VAC @ 60 Hz

Canada - Voltage and Frequency

600 VAC @ 60 Hz

Europe - Voltage and Frequency

400 VAC @ 50 Hz