Troubleshooting, Troubleshooting 44 – MTS Series FEC Environmental Chamber User Manual
Page 44

Routine Maintenance
Series FEC Environmental Chamber Product Manual
Routine Maintenance
44
•
Do not open the chamber immediately after a test to prevent damage or
personal injury caused by high or low temperature.
•
To extend the service life of an environmental chamber, eliminate water
vapor in the cabinet or other residue left in the chamber after test.
•
Do not use corrosive organic solvents when cleaning chamber.
•
During test or after test, do not use any kind of shelter to cover the chamber
so as to avoid the case that the heat or water vapor can not be diffused and
damage the chamber.
Troubleshooting
Troubleshooting
S
YMPTOMS
R
EASON
S
OLUTION
Set Value window flashes and
displays “S.br” .
Measure signal input
exceeds range.
1. Inspect if specification setting of
sensor is correct.
2. Inspect if signal input is open circuit.
Temperature increase speed is
slower than original speed
Parameter setting may be
incorrect.
Inspect if setting of temperature control
parameter is correct.
Temperature rises out of
control
parameter setting is
incorrect.
Inspect if setting of temperature control
parameter is correct.
Temperature fails to mains at
setting temperature
Control parameter is
imperfect or lost.
Start self-adjustment to find out proper
control parameter.