Series FEC Environmental Chamber Product Manual
Introduction
15
Cabinet
The surface of the cabinet is made of painted mild-steel plate, the inner wall of
the cabinet is made of stainless steel making it temperature-resistant. The front
side of the cabinet has a door with an observation window, which you can
observe tests under illumination from the lamp that is installed on the inner wall
of the cabinet. Cooling is accomplished with the use of liquid nitrogen introduced
to the chamber. Heating is accomplished with one electrical heating element. An
electric fan and baffle diffuse the heat for uniform temperatures.
Temperature control
system
The temperature controller unit is integrated into the backside of the cabinet,
which allows you to perform controller adjustment easily. The temperature
control system uses a Eurotherm high-precision temperature controller, and a
solid state relay as an executing element.
Bracket for
environmental
chamber
The bracket can be dismantled or assembled easily.You can move the chamber
back from the test zone of testing machine via rollers on the bottom of the
cabinet.
Note
FEC serial chamber can only be used on MTS frames. When you want to
use it for other purpose, you should be aware of the chamber overturn
risk due to external force.
Motor, Heating and
Refrigeration pipe
Liquid Nitrogen Interface
Bracket
Movable Wheels
Electric Controls
Test Space