beautypg.com

Sterilisers sf/sn, 1 intended use, 3 guidelines for sterilisation – Memmert Universal ovens U, Incubators I, Sterilisers S User Manual

Page 43

background image

D24022 | Date 10/2014

43

Sterilisers SF/SN

8. Sterilisers SF/SN

8.1 Intended use

The SF/SN appliance serves for sterilising medical material through dry heated air at atmos-

pheric pressure.

8.2 Note in accordance with Medical Devices Directive

The product lifetime as intended by the manufacturer is eight years.

8.3 Guidelines for sterilisation

For hot air sterilisation, there are different guidelines on the temperature and sterilisation time

to choose, as well as on packaging the sterilisation load. The values to be chosen depend on

the type and characteristics of the load to be sterilised and on the type of germs to be neutral-

ised. Before beginning sterilisation, make yourself familiar with the sterilisation method laid

down for your application.
Process parameters for hot air sterilisers are temperature and minimum hold time. The follow-

ing process parameters have been defined in recognised standards:

According to WHO: 180 °C with a minimum hold time of 30 min

According to the European Pharmacopoeia: 160 °C with a minimum hold time of 120 min

For the inactivation of endotoxin (pyrogenes), dry heat of at least 180 °C can be applied. For

the depletion of pyrogenic substances, you have to keep a combination of temperature and

time going beyond the requirements of sterilisation.
Inactivation of endotoxin is possible using the following process parameters (data in accord-

ance with ISO 20857:2010):

180 °C with a minimum effective time of 180 min

250 °C with a minimum effective time of 30 min

Caution:

The temperature and time requirements normal for hot air sterilisation do not destroy

endotoxins.

Especially when the appliance is heavily loaded, using these parameters without checking

them will not be sufficient. For safe sterilisation, validation of the individual sterilisation pro-

cess is required. The requirements for the validation of sterilisation by dry heat are e.g. defined

in standard ISO 20857:2010. Also valuable is the "guideline on validation and routine moni-

toring of sterilisation processes using dry heat for medical products" issued by the German

Society for Hospital Hygiene (DGKH).