5 temperature monitoring, Setup, Mechanical temperature limiter (tb) – Memmert Universal ovens U, Incubators I, Sterilisers S User Manual
Page 27

D24022 | Date 10/2014
27
Operation and control
3. Press the confirmation key.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
The display now shows the remaining time in
a large font and the approximate end time in
a smaller font beneath.
TIMER
30
m
04
h
End 13:30 23.11.
4. Now, as described under 5.4.2, set the individual values for temperature, air flap position
etc. which you want the appliance to operate at. For universal ovens UN/UF and incubators
IN/IF, the set values can be changed while the timer elapses. The changes are effective im-
mediately. For sterilisers SN/SF, parameters cannot be changed while the timer elapses.
For universal ovens UN/UF and incubators IN/IF, you can choose if the timer should run
setpoint-dependent or not in the
Setup
– this determines if the timer should not start
until a tolerance band around the set temperature is reached or if it should start right after
activation (see page 38). If the timer runs setpoint-dependent, this is indicated by the
symbol in the timer display.
symbol in the timer display.
When the timer has elapsed, the display shows 00h:00m. All func-
tions (heating etc.) are switched off. If a fan had been active, it will
keep on running for a short safety period. In addition, an acoustic
alarm sounds, which can be turned off by pressing the confirma-
tion key.
To deactivate the timer, open the timer display by pressing the
activation key again and then turning the turn control to reduce
the timer setting until --:-- is displayed. Confirm with the confirma-
tion key.
5.5 Temperature monitoring
The appliance is equipped with a double overtemperature protection (mechanical/electronic)
in accordance with DIN 12 880. This serves to avoid damage to the chamber load and/or ap-
pliance in case of a malfunction:
►
Electronic temperature monitoring (TWW/TWB) (TWB only if equipped with a second tem-
TWW/
TWW/
perature sensor, option A6)
►
mechanical temperature limiter (TB)
TIMER
00
m
00
h
End 13:30 23.11.
TIMER
--
m
--
h
End 9:00 23.11.