beautypg.com

3 loading the appliance, 4 operating the appliance, 1 controlcockpit – Memmert Universal ovens U, Incubators I, Sterilisers S User Manual

Page 23

background image

D24022 | Date 10/2014

23

Operation and control

5.3 Loading the appliance

Warning!

When loading the appliance with an unsuitable load, poisonous

or explosive vapours or gases may be produced. This could cause

the appliance to explode, and persons could be severely injured or

poisoned. The appliance may only be loaded with materials which do

not form any toxic or explosive vapours when heated up and cannot

ignite (see also Intended use on page 8). If there is any doubt as

to the composition of materials, they must not be loaded into the

appliance.

Caution:

Check the chamber load for chemical compatibility with the materials of the appliance

(see page 11).

Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the

load capacity are specified in the technical data overview on page 13.
The chamber must not be

loaded too tightly, so that

proper air circulation in the

working chamber is guaran-

teed. Do not place any

chamber load on the floor,

touching the side walls or

right below the ceiling of the

working chamber (Fig. 13 , see

also the "correct loading"

sticker on the appliance).
In case of improper loading

(chamber loaded too tightly),

the set temperature may

be exceeded or it may take

longer until it is reached.

To achieve the correct heating capacity, the type of slide-in unit used – Grid or Shelf –

must be set in the menu under

SETUP

(see page 38).

5.4 Operating the appliance

5.4.1 ControlCOCKPIT

In manual operation, the desired parameters are entered at the ControlCOCKPIT on the front

of the appliance (Fig. 14 and Fig. 15 ). You can also make basic settings here (menu). Addition-

ally, warning messages are displayed, e.g. if the temperature is exceeded.

Loading – Models 30 - 1060

Open / close the door

E24899

Memmert GmbH + Co. KG | P.O. Box 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: [email protected]

www.memmert.com | www.atmosafe.net

Fig. 13

Correct placement of the chamber load