4 timer mode, 5 type of the slide-in unit (grid or shelf), 6 remote control – Memmert Universal ovens U, Incubators I, Sterilisers S User Manual

Page 38: Remote control, Read only, Write, Write+alarm

background image

38

D24022 | Date 10/2014

Menu functions

7.3.4 Timer Mode

For universal ovens UN/UF and incubators IN/IF, you can choose

whether the digital backwards counter with target time setting

(see page 26) should run setpoint-dependent or not – this

determines if the timer should not start until a tolerance band

of ±3 K around the set temperature is reached ( Fig. 22, B) or if

it should start right after activation (A).

t

Timer

t

Timer

A

B

°C/°F

t

6 K

Fig. 22

Timer Mode

A

Timer independent of setpoint: Timer starts right after activation

Fig. 22

Fig. 22

B

Timer setpoint-dependent: Timer does not start until tolerance band is reached

A

Timer independent of setpoint: Timer starts right after activation

A

Timer independent of setpoint: Timer starts right after activation

For sterilisers SN/SF, the timer is by default setpoint-dependent. To ensure that the required

temperature is maintained for a sufficient period of time, this setting cannot be changed.

If the temperature is no longer within the tolerance band, the sterilisation time is, for

reasons of safety, restarted as soon as the required temperature is reached again. For uni-

versal ovens UN/UF and incubators IN/IF, the timer is in this case interrupted and resumed

as soon as the required temperature is reached again.

7.3.5 Type of the slide-in unit (Grid or Shelf)

Here, you have to set the type of the slide-in unit (grid or shelf)

used. The selection Shelf enables you to adjust the control

function to the different air flow characteristics in the interior

when using optional sliding shelves instead of the grids that are

part of the standard delivery.

7.3.6 Remote control

In the setup entry

Remote control

, you can set whether the

appliance should be controlled via remote control and if

so, in which mode. These settings are available:

Off

Read Only

Write

Write+Alarm

SETUP

Unit

IP address

255.145.136.225

Subnet mask

255.255.0.0

Alarm temp

°C

F

TWW

TWB

Timer mode

Slide-in unit

Grid

Shelf

SETUP

Unit

IP adress

255.145.136.225

Subnet mask

255.255.0.0

°C

°F

Timer mode

Alarm temp

TWW

TWB

Slide-in unit

Grid Shelf

Setup

Balance

Remote Control

Off

Gateway 192.168.5 .1

+30 %

2/2