4 timer mode, 5 type of the slide-in unit (grid or shelf), 6 remote control – Memmert Universal ovens U, Incubators I, Sterilisers S User Manual
Page 38: Remote control, Read only, Write, Write+alarm

38
D24022 | Date 10/2014
Menu functions
7.3.4 Timer Mode
For universal ovens UN/UF and incubators IN/IF, you can choose
whether the digital backwards counter with target time setting
(see page 26) should run setpoint-dependent or not – this
determines if the timer should not start until a tolerance band
of ±3 K around the set temperature is reached ( Fig. 22, B) or if
it should start right after activation (A).
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 22
Timer Mode
A
Timer independent of setpoint: Timer starts right after activation
Fig. 22
Fig. 22
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
A
Timer independent of setpoint: Timer starts right after activation
A
Timer independent of setpoint: Timer starts right after activation
For sterilisers SN/SF, the timer is by default setpoint-dependent. To ensure that the required
temperature is maintained for a sufficient period of time, this setting cannot be changed.
If the temperature is no longer within the tolerance band, the sterilisation time is, for
reasons of safety, restarted as soon as the required temperature is reached again. For uni-
versal ovens UN/UF and incubators IN/IF, the timer is in this case interrupted and resumed
as soon as the required temperature is reached again.
7.3.5 Type of the slide-in unit (Grid or Shelf)
Here, you have to set the type of the slide-in unit (grid or shelf)
used. The selection Shelf enables you to adjust the control
function to the different air flow characteristics in the interior
when using optional sliding shelves instead of the grids that are
part of the standard delivery.
7.3.6 Remote control
In the setup entry
Remote control
, you can set whether the
appliance should be controlled via remote control and if
so, in which mode. These settings are available:
►
Off
►
Read Only
►
Write
►
Write+Alarm
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid
Shelf
SETUP
Unit
IP adress
255.145.136.225
Subnet mask
255.255.0.0
°C
°F
Timer mode
Alarm temp
TWW
TWB
Slide-in unit
Grid Shelf
Setup
Balance
Remote Control
Off
Gateway 192.168.5 .1
+30 %
2/2