6. Malfunctions, warning and error messages

Warning!

After removing covers, live parts may be exposed. You may receive

an electric shock if you touch these parts. Malfunctions requiring

work inside the appliance may only be rectifi ed by electricians. Ob-

serve the separate service manual for this.

Do not try to rectify appliance errors yourself but contact the MEMMERT customer service

department (see page 2) or an authorised service point.
In case of enquiries, please always specify the model and appliance number from the name-

plate (see page 12).

6.1 Warning messages of the temperature monitoring function

Description

Cause

Action

See

Temperature alarm and

"TWW" are displayed

TEMP

TWW Set 190.0

°C

195

.4

°C

The adjustable

undertemperature

/ overtemperature

controller (TWW)

has assumed heat-

ing control.

Increase the difference between

the monitoring and setpoint

temperature – by either in-

creasing the max value of the

temperature monitoring (

MAX

temp

) or decreasing the set-

point temperature.
If the alarm continues: Contact

customer service

page

37

page 2

Temperature alarm and

"TWB" are displayed

TEMP

TWB Set 190.0

°C

195

.4

°C

The electronic

temperature

limiter (TWB) per-

manently switched

off heating.

Deactivate the alarm by pressing

the confirmation key.
Increase the difference between

the monitoring and setpoint

temperature – by either in-

creasing the max value of the

temperature monitoring (

MAX

temp

) or decreasing the set-

point temperature.
If the alarm continues: Contact

customer service

page

37

page 2

Temperature alarm and

TB

are displayed

TEMP

TB

230

.4

°C

The mechanical

temperature lim-

iter (TB) perma-

nently switched

off heating.

Switch off the appliance and

leave to cool down. Contact

customer service and have the

error rectified (e.g. by replacing

the temperature sensor).