5.5 System Monitor
The SYSTEM MONITOR screen supplies important infor-
mation on the Leica ASP300 system status.
After replacing the filter press RESET.
ROTARY VALVE RETRIES:
If a figure higher than 50 is indicated, a preventive maintenance is
recommended.
FILL/DRAIN:
This screen displays the average fill / drain times of all stations
(in seconds).
Short term = average duration of last 5 fills / drains.
Long term = average duration of last 20 fills / drains.
If one of the figures displayed exceeds 250, call Leica Technical
Service.
PUMP AGE:
Indication of hours of service of the pressure/
vacuum pump. After approx. 1,000 hours of
use, have Leica Technical Service carry out a
preventive maintenance of the pump.
CARBON FILTER AGE:
Indication of hours of service of the activated
carbon filter. The carbon filter must be ex-
changed after approx. 4 months of use.
Fig. 33
→
→
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)