Leica ASP300
Fig. 16
→
→
Adding molten paraffin
•
Molten wax may have a maximum temperature of 70 ° C.
•
If filling with molten wax, do not fill above the MAX level (see
Fig. 15
).
•
After a wax bath has been filled, it must be marked as ‘full’ in the
REAGENT STATUS screen.
Software control
•
If the wax baths are filled with molten wax, it is possible to override
the Leica ASP300 melting time calculation.
For that purpose, in supervisor mode, access the
SERVICE FUNCTIONS screen (
Fig. 16
).
•
Press TIME UNTIL WAX MELTED.
•
Enter the required delay time and press OK
to confirm.
If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru-
ment malfunctions may occur.
→
→
5.
Operation
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)