Operation – Leica Biosystems ASP300 User Manual
Page 32

32
Instruction Manual V 1.3 – 12/2003
Fig. 14
→
→
•
Remove the splash protector.
•
Fill the wax bath (
Fig. 15
) with wax pellets up
to the top.
•
Press BACK and REAGENT STATUS to ac-
cess the reagent status table (
Fig. 14
). High-
light the wax bath you just filled and press
SET AS FULL.
•
After about 90 minutes, add another 0.5 kg
of wax pellets.
•
The pellets will take a total of approx. 10
hours to melt.
Important!
During the melting process, the paraf-
fin level MAY NOT drop below the
MIN level (2,
Fig. 15
).
•
Once the paraffin is completely molten, re-
insert the splash protector.
Software control
•
The software calculates when the wax will
be ready for use and selects program start-
ing times accordingly, i.e. the paraffin will
definitely be molten when it is required for
the first paraffin step.
Filling with wax pellets
5.
Operation
Fig. 15
1
2
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)