Error no.: 25

Error no.: 25

Error no.: 25

Error no.: 25

Error no.: 25 Power failure

Power failure

Power failure

Power failure

Power failure

Reason:

Power failure while rotor is in motion

Action:

Open and reclose the lid, restart the centrifuge

Error no.: 36

Error no.: 36

Error no.: 36

Error no.: 36

Error no.: 36 Relay for the frequency converter cannot be released

Relay for the frequency converter cannot be released

Relay for the frequency converter cannot be released

Relay for the frequency converter cannot be released

Relay for the frequency converter cannot be released

Reason:

Defect on the power board

Action:

Call service

Reason:

Centrifuge lid is jamming

Action:

Open the centrifuge lid manually. Grease the lid lock lightly.

If this failure occurs again, call service.

Error no.:50,51

Error no.:50,51

Error no.:50,51

Error no.:50,51

Error no.:50,51

Memory failure

Memory failure

Memory failure

Memory failure

Memory failure

Reason:

Internal or external memory failure

Action:

Restart the unit, if the failure occurs again, call service

Error no.: 55

Error no.: 55

Error no.: 55

Error no.: 55

Error no.: 55 Overspeed

Overspeed

Overspeed

Overspeed

Overspeed

Reason:

Overspeed sensor or engine speed sensor defective

Action:

Call service

Error no.: 60

Error no.: 60

Error no.: 60

Error no.: 60

Error no.: 60 Engine speed sensor signal is missing

Engine speed sensor signal is missing

Engine speed sensor signal is missing

Engine speed sensor signal is missing

Engine speed sensor signal is missing

Reason:

Engine speed sensor defective or parting of a cable at the sensor

Action: