Labnet Z233M-2 High Capacity MicroCentrifuge User Manual
Page 19

19
Error no.: 25
Error no.: 25
Error no.: 25
Error no.: 25
Error no.: 25 Power failure
Power failure
Power failure
Power failure
Power failure
Reason:
Power failure while rotor is in motion
Action:
Open and reclose the lid, restart the centrifuge
Error no.: 36
Error no.: 36
Error no.: 36
Error no.: 36
Error no.: 36 Relay for the frequency converter cannot be released
Relay for the frequency converter cannot be released
Relay for the frequency converter cannot be released
Relay for the frequency converter cannot be released
Relay for the frequency converter cannot be released
Reason:
Defect on the power board
Action:
Call service
Reason:
Centrifuge lid is jamming
Action:
Open the centrifuge lid manually. Grease the lid lock lightly.
If this failure occurs again, call service.
Error no.:50,51
Error no.:50,51
Error no.:50,51
Error no.:50,51
Error no.:50,51
Memory failure
Memory failure
Memory failure
Memory failure
Memory failure
Reason:
Internal or external memory failure
Action:
Restart the unit, if the failure occurs again, call service
Error no.: 55
Error no.: 55
Error no.: 55
Error no.: 55
Error no.: 55 Overspeed
Overspeed
Overspeed
Overspeed
Overspeed
Reason:
Overspeed sensor or engine speed sensor defective
Action:
Call service
Error no.: 60
Error no.: 60
Error no.: 60
Error no.: 60
Error no.: 60 Engine speed sensor signal is missing
Engine speed sensor signal is missing
Engine speed sensor signal is missing
Engine speed sensor signal is missing
Engine speed sensor signal is missing
Reason:
Engine speed sensor defective or parting of a cable at the sensor
Action:
Call service