Function instructions – Klein Tools CL1100 User Manual
Page 4
7
1
5
2
3
6
600V
400V
CATIII
600V
400V
CATIII
PRESIONE
CA CC
V
AT
0
10
20
40
30
9
1.
AC Current (large): < 400A
Features:
2.
AC / DC Current (small): < 2000µA
Features:
MAX/MIN
RANGE
HOLD
J.
Magnetic Back
• Attach instrument to metal for hands-free use.
K.
Bar Graph
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
L.
Temperature Input Jack
• Input jack for K-type thermocouple probe.
M. Temperature Switch
• Move the switch down to measure temperature, and up to
measure all other functions that require standard test leads. Test
leads must be removed before using the temperature function.
N.
Temperature Adjustment
• Remove battery cover.
• Place temperature probe in distilled
water with stirred crushed ice
(a standard for 0°C and 32°F).
• Use a fi ne tip screwdriver to adjust
the screw recessed in the bottom
right corner of the battery cavity as
shown until the display reads 32°F.
FUNCTION INSTRUCTIONS
• Select AC or DC current source.
•
Current above 2000µA will damage instrument.
MAX/MIN
RANGE
HOLD
• Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary).