Klein Tools MM6000/6000N User Manual

Mm6000, Instruction manual

background image

ENGLISH

Instruction

Manual

ENGLISH

MM6000

• true rMs

• auto / Manual

RANGE

• auto HolD

• MaX / Min

• Bar GrapH

• BacKlit DisplaY

• 3-5/6 DiGit

6000 count lcD

• leaD storaGe

4007036

sYMBols useD on lcD

~

ac Measurement

Dc Measurement

-

negative Dc Value

AT

auto range active

O.L.

overload: range exceeded

Apo

auto power-off active

low Battery

HolD

Hold active

MIN

Minimum reading

MAX

Maximum reading

%

Duty cycle Mode

Frequency Mode

V

Voltage Measurement

A

current in amps

Ω

resistance in ohms

Diode test

F

capacitance in Farads

continuity test

relative / Zero Mode

n

nano 10

-9

µ

Micro 10

-6

m

Milli 10

-3

k

Kilo 10

3

M

Mega 10

6

°F

Degrees Fahrenheit

°C

Degrees celsius

A- HOLD

auto-Hold active

LOW

Z

low impedance

electrical speciFications

Dc Voltage Measurement

range

resolution

accuracy

600mV ˜ 600V

0.1mV ˜ 0.1V

± (0.2% + 5 digits)

1000V

1V

± (0.5% + 10 digits)

overload protection:

1000V

input impedance (nominal)

: >10MΩ, <100pF

input impedance (loW Z)

: >3kΩ, <200pF, up to 600V RMS

ac Voltage Measurement

range

resolution

accuracy

40Hz ~ 400Hz

accuracy

400Hz ~ 2kHz

600mV

0.1mV

± (0.75% + 5 digits)

± (2.5% +10 digits)

6V ~ 600V

1mV ~ 0.1V

± (2.0% + 8 digits)

1000V

1V

± (0.75% + 8 digits)

± (2.0% +8 digits)*

Continued on next page.

overload protection:

1000V RMS

input impedance (nominal):

>10MΩ, <100pF

input impedance (loW Z)

: >3kΩ, <200pF, up to 600V RMS

Frequency response:

40 to 2kHz

response:

True RMS

*Accuracy for 400Hz ~ 1kHz range only.
Dc current Measurement

range

resolution

accuracy

600μA ~ 400mA

0.1μA ~ 0.1mA

± (1.0% + 5 digits)

6A ~ 10A

1mA ~ 10mA

± (1.5% + 5 digits)

overload protection: µama: 400mA (F440mA/1000V fuse) / a input: 10A (F11A/1000V fuse)

ac current Measurement

range

resolution

accuracy

40Hz ~ 2kHz

600μA ~ 400mA

0.1μA ~ 0.1mA

± (1.2% + 5 digits)

6A ~ 10A

1mA ~ 10mA

± (1.8% + 5 digits)

overload protection: µama: 400mA (F440mA/1000V fuse) / a input: 10A (F11A/1000V fuse)
Frequency:

40 to 2kHz

response:

True RMS

resistance Measurement

range

resolution

accuracy

600

Ω ˜ 6MΩ

0.1

Ω ˜ 0.001MΩ

± (1.0% + 5 digits)

50M

Ω

0.01M

Ω

± (1.5% + 10 digits)

overload protection:

600V RMS

capacitance Measurement

range

resolution

accuracy

60nF

0.01nF

± (3.5% + 6 digits)

600nF ~ 6000μF

0.1nF ~ 1μF

overload protection:

600V RMS

Frequency Measurement

range

resolution

accuracy

99.99Hz ˜ 399.9kHz

0.01Hz ˜ 0.1kHz

± (0.1% + 3 digits)

overload protection:

600V RMS

sensitivity:

2.5V RMS

Duty cycle Measurement

range

resolution

accuracy

1.0 ˜ 99.0%

0.1%

± (0.2% per kHz + 0.1% + 8 digits)

overload protection:

600V RMS

Sensitivity: < 50V RMS

Frequency range:

0.5Hz to 10kHz

Diode test

overload protection

range

test current

open circuit Voltage

600V RMS

6.0V DC

0.25mA

< 1.8V DC

continuity test

overload protection

open circuit Voltage

threshold (appx.)

600V RMS

< 0.44V

<40

Ω

temperature Measurement

range

resolution

accuracy

-58 ~ 600°F

0.1°F

± (1.0% + 5.4°F)

-50 ~ 600°C

0.1°C

± (1.0% + 3.0°C)

600 ~ 1832°F

1°F

± (1.0% + 5.4°F)

600 ~ 1000°C

1°C

± (1.0% + 3.0°C)

overload protection:

600V RMS

K-type

thermocouple accuracy:

Not specifi ed

Specifi cations subject to change.

WarrantY

www.kleintools.com/warranty

CLEANING

Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.

STORAGE

Remove the batteries when instrument is not in use for a prolonged period
of time. Do not expose to high temperatures or humidity. After a period of
storage in extreme conditions exceeding the limits mentioned in the General
Specifi cations section, allow the instrument to return to normal operating

conditions before using it.

Disposal / recYcle

Do not place equipment and its accessories in the trash.

Items must be properly disposed of in accordance with local

regulations.
Prior to disposal of this product, please contact Klein Tools

for proper disposal options.

custoMer serVice

Do not apply voltage to thermocouple.

• Fahrenheit range: -58° to 1832°F (Celsius range: -50° to 1000°C).

8.

temperature:

Features:

MaX/Min

REL

HolD

testing Fuse:

• When selecting uA or mA/A functions the meter will indicate

lead” if the test leads are not properly connected to the meter.

• Note: This message remaining on the screen after properly

connecting the test leads is an indication that the fuse is open.

Klein tools, inc.

450 Bond Street

Lincolnshire, IL 60069

1-877-775-5346

www.kleintools.com

[email protected]

Rev 04/14 B

139750

7.

Frequency (Hz) / Duty cycle: <400kHz

Features:

HolD

6.

low impedance (loW Z): < 600V

Do not use (loW Z) mode at voltages greater than 600V.

• Press SELECT to choose AC or DC Voltage.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• See AC/DC Voltage instruction.

Capacitance Features:

HolD

safely discharge capacitor before

measurement.

• Reading may take up to 60 seconds

for large capacitors.

auto HolD

1000V

10A

Dwg Name:

MM6000-139750ART Dwg No: 139750

ECO No:

016671 Pkg Dwg Ref: 1650 Rev: B

Color Reference:

CEN-PACK-003-A

MM6000-139750ART-Instructions.indd 1-4

5/5/14 10:19 AM