Klein Tools MM1300 User Manual
Instruction manual, Mm 1300
cuStoMEr SErVIcE
KLEIN tooLS, INc.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
www.kleintools.com
139713 Rev. 03/14 B
ENGLISH
Instruction Manual
• auto /
MaNuaL
raNGE
• auto HoLD
• MaX / MIN
• Bar GrapH
• BacKLIGHt
• LEaD
HoLDEr
• 3-5/6 DIGIt
6000 couNt
LcD
ENGLISH
8. temperature:
Features:
SYMBoLS uSED oN LcD
~
ac Measurement
Dc Measurement
-
Negative Dc Value
AT
auto range active
O.L.
overload: range Exceeded
Apo
auto power-off active
Low Battery
HoLD
Hold active
MIN
Minimum reading
MAX
Maximum reading
%
Duty cycle Mode
Frequency Mode
V
Voltage Measurement
A
current in amps
Ω
resistance in ohms
Diode test
F
capacitance in Farads
continuity test
relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Degrees Fahrenheit
°C
Degrees celsius
a-HoLD
auto-Hold active
ELEctrIcaL SpEcIFIcatIoNS
Dc Voltage Measurement
range
resolution
accuracy
600mV ˜ 600V
0.1mV ˜ 0.1V
± (0.5% + 4 digits)
1000V
1V
± (0.8% + 10 digits)
overload protection:
1000V
Input Impedance (Nominal): 10MΩ, <100pF
ac Voltage Measurement
range
resolution
accuracy
600mV
0.1mV
± (0.75% + 5 digits)
6V ~ 600V
1mV ~ 0.1V
1000V
1V
± (1.0% + 8 digits)
overload protection:
1000V RMS
Input Impedance (Nominal): 10MΩ, <100pF
Frequency response:
40 to 400 Hz
response:
Averaging
Dc current Measurement
range
resolution
accuracy
600µA ~ 400mA
0.1µA ~ 0.1mA
± (1.0% + 5 digits)
6A ~ 10A
1mA ~ 10mA
± (1.5% + 5 digits)
overload protection:
•
µama:
F440mA/1000V fuse
•
a Input:
F11A / 1000V fuse
ac current Measurement
range
resolution
accuracy
600µA ~ 400mA
0.1µA ~ 0.1mA
± (1.5% + 5 digits)
6A ~ 10A
1mA ~ 10mA
± (2.0% + 5 digits)
overload protection: µama:
F440mA/1000V fuse /
a Input:
F11A / 1000V fuse
Frequency:
40 to 400Hz
response:
Averaging
resistance Measurement
range
resolution
accuracy
600
Ω ˜ 6MΩ
0.1
Ω ˜ 0.001MΩ
± (1.0% + 5 digits)
60M
Ω
0.01M
Ω
± (1.5% + 10 digits)
overload protection:
600V RMS
capacitance Measurement
range
resolution
accuracy
60nF
0.01nF
± (3.5% + 6 digits)
600nF ~ 6000µF
0.1nF ~ 1µF
overload protection:
600V RMS
Frequency Measurement
range
resolution
accuracy
99.99Hz ˜ 499.9kHz
0.01Hz ˜ 0.1kHz
± (0.1% + 3 digits)
overload protection:
600V RMS
Sensitivity:
>0.7V RMS
Duty cycle Measurement
range
resolution
accuracy
0.1 ˜ 99.9%
0.1%
± (0.2% per kHz + 0.1% + 5 digits)
overload protection:
600V RMS
Sensitivity:
< 50V
Frequency range:
<10kHz
Diode test
overload protection
range
test current
open circuit Voltage
600V RMS
6.0V
Appx. 0.25mA
< 1.8V DC (typ.)
continuity test
overload protection
open circuit Voltage
threshold (appx.)
600V RMS
< 0.44V
<40
Ω
temperature Measurement
range
resolution
accuracy
-58 ~ 600°F
0.1°F
± (1.0% + 5.4°F)
-50 ~ 600°C
0.1°C
± (1.0% + 3.0°C)
600 ~ 1832°F
1°F
± (1.0% + 5.4°F)
-600 ~ 1000°C
1°C
± (1.0% + 3.0°C)
overload protection:
600V RMS
thermocouple accuracy:
Not specified
Specifi cations subject to change
7. capacitance:
< 6000µF
Features:
HoLD
Safely discharge capacitor before measurement.
• Reading may take up to 60 seconds for large capacitors.
Do Not apply voltage to thermocouple.
• -58° to 1832°F (-50° to 1000°C)
testing Fuses:
• When selecting µA or mA/A functions the meter will indicate
“LEad” if the test leads are not properly connected to the meter.
• Note: This message remaining on the screen after properly
connecting the test leads is an indication that the fuse is open.
MaX/MIN
HoLD
rEL
a- HoLD
6.
Frequency (Hz) / Duty cycle <500kHz
Features:
HoLD
WarraNtY
www.kleintools.com/warranty
cLEaNING
Clean the instrument by using a damp cloth. Do not use abrasive cleaners
or solvents.
Take care to keep the sensor lens clean at all times. Clean the lens using a soft
cloth or cotton swab with water or rubbing alcohol only, and allow the lens to
dry before use.
StoraGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISpoSaL / rEcYcLE
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
1000V
10A
Dwg Name:
MM1300-139713ART Dwg No: 139713
ECO No:
16082 Pkg Dwg Ref: 1650 Rev: B
Color Reference:
N/A
PRESS
PRESS
MM
1300
MM1300 Instruction Manuals - MASTER.indd 1-4
3/12/2014 10:16:56 AM