KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Programmable Option– RS232 Commands and Responses

(Continued)

mode prgm

sets pump in program mode. Response :

number n

sets number of steps in program. n = 1 - 8

step n

sets step number to be programmed. n = 1 - 8

important that step number be set before entering program settings

time xx:xx:xx

sets time duration of step to be programmed. hr:min:sec

travel i (w)

sets direction to infusion ( withdrawal )

rateb nnnnn uuu

sets step beginning rate

where nnnnn is . , 0 to 9

where uuu are units µlm , µlh, mlm, mlh.

Note: if set rate is out of range then response is NA; rate is set

a zero

ratef nnnnn uuu

sets step finish rate

where nnnnn is . , 0 to 9

where uuu are units µlm, µlh, mlm, mlh.

Note: if set rate is out of range then response is NA; rate is set

a zero

portout p

sets status of output pins 1 and 6

where p is

HH ( 1=high, 6=high )

HL

( 1=high, 6=low )

LH

( 1=low, 6=high )

LL

( 1=low, 6=low)

pause y or n

sets status of pause

y = active

n = inactive

loop y or n

sets loop status

y = yes

n = no

loopto n

set step number to loop to where n = 1 to 8

loopcnt b

sets number of loops to be repeated

where b = 1 to 100

save

saves step settings important that each step is saved

done

saves all programmed steps

important that "done" is entered after all steps saved

wait

stops pump (pauses), but can be restarted

continue

restarts pump after "wait" command, program continues

nextstep

causes program to jump to the next step

mode?

query mode. Response PGM

activestep?

queries step running

response: n where n = 1 to 8