Kds model 410 series syringe pumps, Programmable option–features – KD Scientific 410 User Manual
Page 22

20
KDS Model 410 Series Syringe Pumps
Programmable Option–Features
PARAMETERS WHICH CAN BE PROGRAMMED IN EACH STEP
Time duration
Infuse or withdraw
Start flowrate
End flowrate
TTL output settings
Pause, wait for TTL input actuation
Loops to repeat previous steps
To simplify the programming, previously programmed settings are stored in non-volatile
memory and are displayed when Program Mode is selected. Whenever possible, options
are displayed with the "active" option flashing. If flashing, this parameter can be selected
or changed.
The pump can be programmed by first selecting M
MO
OD
DE
E on the mainline menu and then
selecting Program (PRGM).
DISPLAY AND PROGRAMMING SEQUENCE
After selecting Program Mode, display reads:
Table
Dia
Step
Mode
DIA and TABLE
If the pump was previously used in Program Mode the pump will be initialized in
Program Mode when it is switched on. For convenience, should a syringe change be
required, it is possible to enter a new syringe size, either from the stored Table or DIA,
without leaving the Program Mode.
If a syringe size change is made however, this will change all program values to
defaults and will require reprogramming.
It is possible to review the syringe size in "Dia" or "Table" without changing the pro-
grammed settings.
In "Table" review the settings but select "QUIT", do not enter a diameter.
In "Dia" the settings will not change if there is no change to the diameter entered.
MODE
Mode selection reverts back to other pump operations.
STEP
Step selection starts the programming sequence.