10

KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

SYRINGE LOADING (Continued)

5.

Place the syringe in the holder and place the “V” Clamp

(8)

inverted, over

the barrel of the syringe, and replace the Shoulder Screws

(1 & 2)

. Tighten the

Shoulder screws. Make sure the flange of the syringe is resting against the Syringe

holder block, between the clamp and the Syringe Block, and the Plunger Flange is

between the Pusher Bracket and the Pusher Block.

6.

Slide the Flange Bracket towards the Syringe Flange and tighten the two “T”

Knobs

(3 & 4)

.

7.

Slide the Plunger Bracket towards the Syringe Plunger and the Pusher Block and

tighten the two “T” knobs

(5 & 6)

.

8.

Engage the Pusher Block by pulling the Plunger Knob

(7)

out and rotate the

knob 90˚ clockwise.

To simplify syringe loading, the pusher block can be disengaged from the leadscrew by turn-

ing the knob

(7)

and manually moved along the guide rods. Alternately, the Fast forward,

Fast reverse feature can be used (press Run and respective Arrow key simultaneously).

KDS410 WITHDRAWAL OPERATION

For withdrawal, or refill operation, the syringe plunger and barrel flange must be secured by

the supplied brackets.