KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

RS232 COMMANDS (continued)

error?

Errors 8 - 15 only occur if using pressure switch.

8

= overpressure

9

= serial error overpressure

10 = stall + overpressure

11 = stall + serial overrun + overpressure

12 = serial overrun + overpressure

13 = serial error + serial overrun + overpressure

14 = stall + serial overrun + overpressure

15 = stall + serial error + serial overrun + overpressure

Note: sending query also clears all errors.

A serial error indicates a command that is too long for the input buffer.

A serial overrun indicates that a command has been sent before the

prior command has been processed.

A stall error indicated that a stall condition has occurred.

Essential that after each command you must wait for the prompt (indicating that the

command has been executed) before sending the next command.

Also, the prompt will indicate that the command has been executed successfully, or not.

prom?

Queries software version response is number 2100.0xx or 2101.0xx

or similar.