Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual
Page 20

18
KDS Model 410 Series Syringe Pumps
KDS Model 410 Series Syringe Pumps
Operating Instructions
(Continued)
RS232 COMMANDS (continued)
error?
Errors 8 - 15 only occur if using pressure switch.
8
= overpressure
9
= serial error overpressure
10 = stall + overpressure
11 = stall + serial overrun + overpressure
12 = serial overrun + overpressure
13 = serial error + serial overrun + overpressure
14 = stall + serial overrun + overpressure
15 = stall + serial error + serial overrun + overpressure
Note: sending query also clears all errors.
A serial error indicates a command that is too long for the input buffer.
A serial overrun indicates that a command has been sent before the
prior command has been processed.
A stall error indicated that a stall condition has occurred.
Essential that after each command you must wait for the prompt (indicating that the
command has been executed) before sending the next command.
Also, the prompt will indicate that the command has been executed successfully, or not.
prom?
Queries software version response is number 2100.0xx or 2101.0xx
or similar.