Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual
Page 17

15
KDS Model 410 Series Syringe Pumps
KDS Model 410 Series Syringe Pumps
Operating Instructions
(Continued)
RS232 COMMANDS
RS232 is used for remote computer control of up to 100 pumps, identified with an address
from 0 to 99 and set to the correct baud rate. Each pump can be controlled either from the
keypad or via RS232 at all times, but the pump can only respond to one command at a
time. When under RS232 control the display reads "REMOTE". All settings made via RS232
are stored in non-volatile memory.
To move the pump from Remote (RS232) to keypad control press select.
After each command is received and executed by the pump the pump responds with a mes-
sage and a prompt.
Commands and Responses
Commands are not case sensitive.
After each command is received and executed the pump responds with prompt sequence:
abbreviations:
= carriage return
= line feed
= space
= no space
a) Commands:
Address
b) Response:
For example:
Query pump 2 for withdrawal flow rate ( which is 0.2 ml/minute )
command:
2
response:
0.2
Query a single pump for infusion flow rate ( which is 0.2 ml/minute )
command:
ratei?
response:
0.2