Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual

Page 14

background image

12

KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING

While the pump continues to run press select to return to the main menu. Scroll through the

menu and select Volume to display the set dispense volume.

a)

No volume change

Press select. The display returns to the incrementing display volume.

b)

Volume change

1)

Make the changes with the numerical keypad and enter.

2)

The display moves to RATE, permitting a change if required. Use the numerical

keypad and enter to make changes (or leave unchanged). The pump

immediately changes to the new flowrate, if changed, and the volume

continues to increment, uninterrupted by the review process, to new target

dispense volume when it will stop automatically.

Note: If the volume is changed to a volume smaller than the volume already accumu-

lated then the pump will stop as soon as the new, smaller target volume is entered.

MODE SELECTION

Mode selection is available only on infusion/withdrawal models.

Select MODE from the main menu and then scroll through the options displayed and select

the mode required.

Possible modes are: infusion, withdrawal, infusion / withdrawal,

withdrawal / infusion, continuous

Note: for bi-directional modes a volume is required.

infusion

Pump infuses at the set rate and stops automatically when the target volume is reached.

The pump can be manually stopped and restarted at any time, that is, the dispense is

paused and, when restarted will continue to the set dispense volume.

withdrawal

Pump withdraws at the set rate to the set volume.

infusion / withdrawal

The pump first infuses and when the target volume is reached it immediately changes

direction and withdraws. The volume settings for infusion and withdrawal can be differ-

ent, as can the infusion and withdrawal flow rates.

withdrawal / infusion

The pump runs first in the withdrawal direction and then automatically changes to the infu-

sion direction. Different settings of rate and volume for withdraw and infuse are permitted.

continuous

The pump first infuses and then withdraws, and then cycles continuously. Only one vol-

ume setting for infusion and withdrawal is permitted. If the pump is matched to a valve,

which is actuated by a TTL pulse from the pump, this mode can be used to infuse and

then refill the pump for continuous operation.

The menu now prompts for volume and rate settings relevant to the mode selected.