KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

SYRINGE LOADING

(1)

“T” Knob for securing

the adjustable Clamp Rod

(7)

Plunger Knob

(3 & 4)

“T” Knob

(2 places) for

adjusting the

Flange Bracket

(5 & 6)

“T” Knob

(2 places) for

adjusting the

Plunger Knob

(2)

Adjustable Clamp

Flange Bracket

Pusher

Block

Clamping for Plastic & Glass Syringes

Using the Adjustable Clamp for Plastic & Glass Syringes

1.

Loosen “T” knob

(1)

allows the user to lift and rotate the adjustable clamp

(2)

out

of the way.

2.

Loosen “T” knobs

(3 & 4)

and slide the Flange Bracket towards the Pusher Block.

3.

Loosen “T” knobs

(5 & 6)

and move the Plunger Bracket towards the Syringe

Holder Block.

4.

Rotate the Pusher Block Knob

(7)

until it snaps into place. This will disengage the

Pusher Block and allow the user to position the block to accept the syringe.

5.

Place the syringe in the holder, making sure the flange is resting against the Syringe

Holder Block and the plunger is against the Pusher Block. Make sure the plunger

flange rests between the bracket and the pusher block, and the syringe flange

rests between the flange bracket and the syringe holder block for proper operation.

6.

Rotate the Adjustable Clamp

(2)

until it rests on the Syringe barrel.

7.

While pushing down on the Adjustable Clamp

(2)

, tighten the “T” knob

(1)

.

Syringe

Holder Block

Plunger Bracket

KDS Model 410 Series Syringe Pumps