Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual
Page 10

8
KDS Model 410 Series Syringe Pumps
Operating Instructions
(Continued)
SYRINGE LOADING
(1)
“T” Knob for securing
the adjustable Clamp Rod
(7)
Plunger Knob
(3 & 4)
“T” Knob
(2 places) for
adjusting the
Flange Bracket
(5 & 6)
“T” Knob
(2 places) for
adjusting the
Plunger Knob
(2)
Adjustable Clamp
Flange Bracket
Pusher
Block
Clamping for Plastic & Glass Syringes
Using the Adjustable Clamp for Plastic & Glass Syringes
1.
Loosen “T” knob
(1)
allows the user to lift and rotate the adjustable clamp
(2)
out
of the way.
2.
Loosen “T” knobs
(3 & 4)
and slide the Flange Bracket towards the Pusher Block.
3.
Loosen “T” knobs
(5 & 6)
and move the Plunger Bracket towards the Syringe
Holder Block.
4.
Rotate the Pusher Block Knob
(7)
until it snaps into place. This will disengage the
Pusher Block and allow the user to position the block to accept the syringe.
5.
Place the syringe in the holder, making sure the flange is resting against the Syringe
Holder Block and the plunger is against the Pusher Block. Make sure the plunger
flange rests between the bracket and the pusher block, and the syringe flange
rests between the flange bracket and the syringe holder block for proper operation.
6.
Rotate the Adjustable Clamp
(2)
until it rests on the Syringe barrel.
7.
While pushing down on the Adjustable Clamp
(2)
, tighten the “T” knob
(1)
.
Syringe
Holder Block
Plunger Bracket
KDS Model 410 Series Syringe Pumps