KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

RS232 COMMANDS (continued)

ratei nnnnn u/u

Sets infusion rate.

Where is nnnnn is ., 0 to 9, and u/u are units.

Units required but if not specified then defaults to automatic setting

based on syringe diameter.

ratew nnnnn u/u

Sets withdrawal rate.

voli nnnnn uu

Sets infusion target volume.

Where nnnnn is ., 0 to 9

uu are units µl or ml

If units not specified then defaults to automatic setting.

volw nnnnn uu

Sets withdrawal target volume.

voli?

Queries volume setting.

Response: nnnnnuu

Where nnnnn is ·,0 to 9

uu are units µl and ml

mode I

Sets mode to infusion.

mode w

Sets mode to withdrawal.

mode I/w

Sets mode to infusion / withdrawal.

mode w/I

Sets mode to withdrawal / infusion.

mode con

Sets mode to continuous.

Note: A dispense volume must be entered before selecting I/W, W/I

and CON modes.

mode?

Query mode.

Response I, W, I/W, W/I, CON

dir rev

Changes direction of running pump. Available only in Infusion and

Withdrawal modes.

dir?

Query direction.

Response I (infusion) or W (withdrawal).

Not applicable in infusion only models.

error?

Command returns values from 0 to 7 as listed below:

0 = no errors

1 = serial error

2 = stall

3 = stall + serial error

4 = serial overrun

5 = serial error + serial overrun

6 = stall + serial overrun

7 = stall + serial error + serial overrun