Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual
Page 19

17
KDS Model 410 Series Syringe Pumps
KDS Model 410 Series Syringe Pumps
Operating Instructions
(Continued)
RS232 COMMANDS (continued)
ratei nnnnn u/u
Sets infusion rate.
Where is nnnnn is ., 0 to 9, and u/u are units.
Units required but if not specified then defaults to automatic setting
based on syringe diameter.
ratew nnnnn u/u
Sets withdrawal rate.
voli nnnnn uu
Sets infusion target volume.
Where nnnnn is ., 0 to 9
uu are units µl or ml
If units not specified then defaults to automatic setting.
volw nnnnn uu
Sets withdrawal target volume.
voli?
Queries volume setting.
Response: nnnnn
Where nnnnn is ·,0 to 9
uu are units µl and ml
mode I
Sets mode to infusion.
mode w
Sets mode to withdrawal.
mode I/w
Sets mode to infusion / withdrawal.
mode w/I
Sets mode to withdrawal / infusion.
mode con
Sets mode to continuous.
Note: A dispense volume must be entered before selecting I/W, W/I
and CON modes.
mode?
Query mode.
Response I, W, I/W, W/I, CON
dir rev
Changes direction of running pump. Available only in Infusion and
Withdrawal modes.
dir?
Query direction.
Response I (infusion) or W (withdrawal).
Not applicable in infusion only models.
error?
Command returns values from 0 to 7 as listed below:
0 = no errors
1 = serial error
2 = stall
3 = stall + serial error
4 = serial overrun
5 = serial error + serial overrun
6 = stall + serial overrun
7 = stall + serial error + serial overrun