11.4. Status messages / error messages

The instrument sends data (including error messages) only when the computer sends a query.


Status messages

Description

00 MANUAL STOP

Presto in "OFF" state.

01 MANUAL START
or
01 MANUAL START,DEGASING

Presto in keypad control mode
or
Presto in keypad control mode and degasifying active

02 REMOTE STOP

Presto in "r OFF" state.

03 REMOTE START
or
03 REMOTE START,DEGASING

Presto

®

in remote control mode.

or
Presto in remote control mode and degasifying active





Error messages

Description

-02 REFRIGERATOR ALARM

Defect temperature sensor in refrigeration circuit.

-03 EXCESS TEMPERATURE WARNING

High temperature warning "

".

-04 LOW TEMPERATURE WARNING

Low temperature warning "

".

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or
interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 50 K.

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE MEASUREMENT

ALARM

Error in A/D converter.

-13 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

Value lies outside the adjusted range for the high
and low temperature warning limits. But value is
stored.

-14 TEMPERATURE/LEVEL ALARM

Safety temperature alarm
Low level alarm.