background image

Preparations

26


Diagram 1:

JULABO Thermal oils

Change in volume in dependence on the temperature of bath liquid.

Example A: Filling quantity at ambient temperature

5 liters

Intended working temperature range

-40 °C bis +250 °C

Change in volume caused by temperature 1,75 liters

Here an external expansion vessel is required (Order no.: 8 970 830)

Example B: Filling quantity at ambient temperature

5 liters

Intended working temperature range

+20 °C bis +200 °C

Change in volume caused by temperature 1,2 liters

Here no external expansion vessel is required

(see table „Filling volume“ Page 31).